A comparison between a hinged plate and screw system and a conventional tension-band plate and screw system used for correction of an angular deformity of the lower limb: an animal study

Abstract

Background

The purpose of the animal study is to introduce a newly designed hinged plate and screw system for correction of angular deformities of the lower limbs. The technique was compared with the use of a conventional tension-band plate and screw system.

Methods

This is a randomized controlled animal trial. Eight 3-month-old Bama miniature pigs were used to establish animal models. In each animal, one leg was randomly allocated into study group and another leg into control group. Legs of the study group were corrected with a hinged plate and screw system, and legs of the control group were corrected with a conventional tension-band plate and screw system. The corrective rates of medial slope angle, medial proximal tibial angle, and angle of the two arms of the hinged plate were measured. Residual stress on the implants was also evaluated. A P < 0.05 was statistical significant.

Results

At the final measurements of 18 weeks, the mean corrective rates of medial slope angle, medial proximal tibial angle, and angle of the two arms of the study group were 0.71°/week, 0.85°/week, and 2.18°/week, respectively; the data in the control group were 0.84°/week, 0.89°/week, and 2.13°/week, respectively. No significant difference was found between the groups regarding the mean corrective rates of the angles (P < 0.05). The mean residual stress in the study group was 643.35 MPa, and measurement in the control group was 1,273.63 MPa, with a significant difference (P < 0.05).

Conclusions

Compared to the conventional tension-band plate and screw system, the hinged plate and screw system may be more reliable for correction of angular deformities of the lower limb.

Introduction

Angular deformities of the lower limb may be congenital or acquired in children, and the treatments are still challenging problems [1]. Temporary hemiepiphysiodesis with an implant to arrest the growth of the bone is an effective method to correct the deformities, but appropriate implant selection is still controversial [1-3].

In 1949, Blount et al. [2] developed the epiphyseal stapling technique to correct angular deformities by controlling bone growth. The technique is effective, but systematical failure often occurs due to weak support of the staples. In 2006, Stevens [3] developed a tension-band plate and screw system to reduce fixation failure. The system is stronger and now commonly used in clinical practices, resulting in success rates of 78.5% to 94.1% [4-10]. However, that straight plate rarely matches the convex contour of the physis and metaphysic. The poorly mounted implant system is relatively unstable and may produce nonuniform stress distribution on the growth plate and stress concentration on the screws [11]. Thus, fixation failure may happen, especially in patients with abnormal physis [12]. In a series of 31 cases, Schroerlucke reported 26% of fixation failure in the treatment of angular deformities of the lower limb [13]. In addition, the middle portion of the straight plate may produce excessive compression on the periosteum and perichondrium over the convex surface of the physis and epiphysis, resulting in unpredictable complications [14].

The purpose of the study was to introduce a hinged plate and screw system for correction of angular deformities of the lower limb in animal models (Figures 1 and 2). We also compared the use of the implant with the use of the conventional tension-band plate and screw system in the experimental cohort study.

Figure 1
figure 1

The newly designed hinged plate and screw system. (A) The plate has a hinge and two arms and two 2 mm diameter screws. (B) Maximal angle of the two arms. (C) Minimal angle of the two arms.

Figure 2
figure 2

Sketch map of hemiepiphysiodesis. (A) The hinged plate almost completely matches the contour of the bone. (B) The straight conventional plate does not match the bone surface.

Materials and methods

From August 2013 to January 2014, nine 3-month-old Bama miniature pigs were used to establish animal models. All the animals were male. In each animal, one researcher randomly allocated one leg into study group and another leg into control group. The random method was flipping a coin. This study was approved by the Institutional Review Board/Ethics Committee of Xin-Hua Hospital.

Anesthesia and operative techniques

The animals were fasted for 24 hours prior to surgery and were sedated with an intramuscular injection of diazepam and 3% pentobarbital sodium (2 ml/kg). Operations were performed under general anesthesia by intravenous propofol (10 mg/kg/h) and fentanyl (25 mg/kg/h). Cefazolin Sodium (Sinopharm Chemical Reagent Co., Ltd, Shanghai, China) was injected prior to skin incision and every morning for the following 3 days. Postoperative pain was managed with tramadol.

Study group

The animal was placed in the supine position on a radiolucent table. A 2–3-cm-long incision was made on the medial aspect of the proximal tibia (Figure 3). Dissection went deep to the tibia with preservation of the periosteum. In the mid-coronal plane, a 1-mm Keith needle was drilled into the physis. That procedure was performed under guidance of fluoroscope. A hinged plate (HP group, Shanghai Puwei Medical instrument Co., Ltd., Shanghai, China) was placed extra-periosteally and mounted over the needle to center it upon the physis. The cortex was predrilled with a 1.8-mm drill bit so as to reduce stress during screw insertion [3,4]. The screws should be placed parallel or slightly convergent (0° to 20°) without encroaching on the growth plate cartilage. The two screws were tightened alternately.

Figure 3
figure 3

Intraoperative photos. (A) A hinged plate and screw system is implanted. (B) No obvious prominence observed after wound closure.

Control group

As described by Stevens [3], surgical procedures were performed using a tension-band plate and screw system (CP group, Shanghai Puwei Medical instrument Co., Ltd., Shanghai, China).

Postoperative managements

Immediately after operation, we assessed the medial slope angle (MSA, the angle formed between the line vertical to the anatomical axis of the tibia and the line tangent to the medial tibial plateau), the medial proximal tibial angle (MPTA, the angle formed between the anatomical axis of the tibia and the line drawn between the medial and lateral corners of the tibial plateau), and the angle of the two arms (ATA) on anterioposterior fluoroscopic images. Radiographs were also obtained at the end of 6, 12, and 18 weeks postoperatively (Figure 4). Immediately after the last radiography, the animal was scarified and the implant was removed. The epiphysis and metaphysis with the periosteal and perichondral beneath the implant, as well as the adjacent epiphyseal growth plate and metaphysic, were harvested. The sample was fixed in 10% formalin and embedded in methyl methacrylate. In the frontal plane, the sample was cut into 7-mm-thick sections, and the sections were stained with safranin-O/fast green (Sinopharm Chemical Reagent Co., Ltd, Shanghai, China).

Figure 4
figure 4

Measurements on standard anterioposterior radiograph of the knee joint. MSA, medial slope angle; MPTA, medial proximal tibial angle; and ATA, angle of the two arms.

Measurements of the samples

All measurements were conducted in a blinded fashion. The measurements were performed by two of the authors independently, and the mean value of the two measurements was used for analysis. The agreement between the two authors was evaluated with the Spearman correlation coefficient for interrater agreement and intraclass correlation coefficient.

Efficacy of hemiepiphysiodesis was assessed with the method described by Kanellopoulos et al. [15]. On anterioposterior radiograph, we measured three angles, i.e., the MSA, MPTA, and ATA (Figure 5).

Figure 5
figure 5

Standard anterioposterior radiographs of the tibia. (A) Immediately after correction with a hinged plate and screw system. (B) Eighteen weeks after operation. (C) Immediately after correction with a conventional plate and screw system. (D) Eighteen weeks after operation.

Residual stress (remaining stress of the sample after implant removal) at the rim of the metaphyseal screw hole was measured using X-ray diffraction (XRD, Bruker D8 Advance, Karlsruhe, Germany) [16,17]. Reflection of titanium alloy was set at 2theta = 70.631. Lattice strains were measured at six positions (psi values: 0.00, 11.5, 16.4, 20.3, 23.6, and 26.6). Residual stress was calculated using Leptos (Version 7.0, Bruker-AXS, Karlsruhe, Germany) according to the strains at six positions [16,17]. For histological study, the width of perichondrial tissue between the center of the plate and physis was measured on the 40 × magnified image of each section.

Analysis

Statistical analysis was made using the paired t-test for comparing the values of all measurement parameters between the groups at the same interval, and 95% confidence intervals for pairwise differences were calculated. The level of statistical significance was set at P < 0.05.

Results

One sample of the control group was excluded because of fixation failure at 6 weeks postoperatively. That animal was excluded. One animal had a superficial infection and recovered after 1-week application of antibiotics.

In the study group, the angles of MSA, MPTA, and ATA at 6, 12, and 18 weeks were shown in Table 1. The corrective rates of the angles of MSA, MPTA, and ATA were 0.71°/week, 0.85°/week, and 2.18°/week, respectively. The residual stress at the rim of metaphyseal screw hole was 643.35 ± 47.24 MPa (Table 1). The periosteum and perichondrium (stained bluish green) were deformed but were still intact (Figure 6). The mean width of perichondrial tissue was 155.38 ± 8.34 μm.

Table 1 Demographics and measurement values of study group
Figure 6
figure 6

Images of histological sections. The bluish green area shows periosteal and perichondral (Safranin-O/fast green stain; ×40). (A) A section obtained from the tibia of a 3-month-old Bama miniature pig. The pig used for comparison did not have an operation. (B) The periosteal and perichondral beneath the hinged plate are intact after 18 weeks. (C) The periosteal and perichondral beneath the conventional tension-band plate are almost vanished after 18 weeks.

In the control group, the angles of MSA, MPTA, and ATA at 6, 12, and 18 weeks were shown in Table 2. The corrective rates of the angles of MSA, MPTA, and ATA were 0.84°/week, 0.89°/week, and 2.13°/week, respectively. The residual stress at the rim of metaphyseal screw hole was 1,273.63 ± 132.28 MPa (Table 2). The periosteum and perichondrium (stained bluish green) were disorganized and damaged (Figure 6). The mean width of perichondrial tissue was 61.83 ± 7.56 μm.

Table 2 Demographics and measurement values of control group

In comparison, there was no significant difference in any of the baseline angles between the groups. There were significant differences regarding the angle of ATA at 6 weeks (P = 0.0223) and 12 weeks (P = 0.0180), but there was no significant difference at 18 weeks (P = 0.6147). We found no significant difference regarding the angle of MPTA at 0 week (P = 0.3935,), 6 weeks (P = 0.1255), 12 weeks (P = 0.8299), and 18 weeks (P = 0.5565) intervals. No significant difference existed regarding the angle of MSA at 0 week (P = 0.3121), 6 weeks (P = 0.0879), 12 weeks (P = 0.1935), and 18 weeks (P = 0.3635) interval. We found significant differences in residual stress (P = 0.001) and width of perichondrial tissue between the center of the plate and physis (P = 0.0001). The Spearman’s correlation coefficients for interrater agreement and intraclass correlation coefficient (ICC) were 0.87 and 0.92, respectively.

Discussion

Angular deformities of the lower limb are common clinical problems encountered in pediatric orthopedic practices. The disorders are a part of the constellation of developmental orthopedic diseases affecting young patients. The deformities can be either valgus or varus and most commonly affect the knee joint [1-3].

The rational behind the use of staples and conventional tension-band plate and screw system is based on Stevens et al.’s theory [3]. Our newly designed system is also based on the theory. The system can correct angular deformities by arresting growth of the other side of the bone. The plate has two arms and a built-in hinge. Based on the previous studies, we designed the rotation of the two arms ranged from 155° to 170° to better fit the contour of the physis in all stages of angular correction. That automatic change can also disperse repeated stress on the surface of periosteum and perichondrium during walking. Therefore, the risk of fixation failure can be reduced. Because of a better match between the implant and bone surface, residual stress at the rim of metaphyseal screw hole was lower after implant removal.

Our histological study showed that the periosteum and perichondrium beneath the hinged plate were intact, but the structures beneath the conventional plate were almost vanished. Aykut et al. [14] verified that healthy periosteal and perichondral were critical for preserving vigor of the growth plate. Our plate with a compression-free hinge may be more appropriate for protecting the growth plate.

Compare to the conventional tension-band plate and screw system, our system has advantages. The self-changeable arc of the two arms produces uniform stress distribution on the growth plate, which may protect the growth plate and perichondrial ring from compressive injuries. The self-changeable arc may also reduce stress concentration on the screws, which decreases the risk of implant failure. The disadvantage of our system is that there is a prominence owing to the bulged hinge, but the morbidity is minimal. Our system can be an alternative to the conventional tension-band plate and screw system, especially in obese patients and patients with poor bone quality such as Blount’s disease. It seems no contraindication compare to the conventional technique.

Our study has limitations. The small sample size affects the reliability of the results because it leads to a high variability. Our assessments on stress distribution and residual stress are based on the changes on images rather than direct measurements, which may not accurately reflect the inherent stress.

Conclusion

Compared to the conventional tension-band plate and screw system, the use of the hinged plate and screw system may be a more reliable technique with minimal complications for correction of angular deformities of the lower limb.

Abbreviations

CP:

Conventional plate

HP:

Hinged plate

MSA:

Medial slope angle

MPTA:

Medial proximal tibial angle

ATA:

Angle of the two arms

References

  1. Paley D. Growth plate consideration. In: Paley D, editor. Principles of deformity correction. New York: Springer- Verlag Berlin Heidelberg; 2005. p. 695–716.

    Google Scholar 

  2. Blount WP, Clarke GR. Control of bone growth by epiphyseal stapling: a preliminary report. J Bone Joint Surg Am. 1949;31:464–78.

    Google Scholar 

  3. Stevens PM. Guided growth: 1933 to the present. Strat Traum Limb Recon. 2006;1:29–35.

    Article  Google Scholar 

  4. Stevens PM. Guided growth for angular correction: a preliminary series using a tension band plate. J Pediatr Orthop. 2007;27:253–9.

    Article  PubMed  Google Scholar 

  5. Boero S, Michelis MB, Riganti S. Use of the eight-Plate for angular correction of knee deformities due to idiopathic and pathologic physis: initiating treatment according to etiology. J Child Orthop. 2011;5:209–16.

    Article  PubMed Central  PubMed  Google Scholar 

  6. Jelinek EM, Bittersohl B, Martiny F, Scharfstädt A, Krauspe R, Westhoff B. The 8-plate versus physeal stapling for temporary hemiepiphyseodesis correcting genu valgum and genu varum: a retrospective analysis of thirty five patients. Int Orthop. 2012;36:599–605.

    Article  PubMed Central  PubMed  Google Scholar 

  7. Burghardt RD, Herzenberg JE. Temporary hemiepiphysiodesis with the eight-Plate for angular deformities: mid-term results. J Orthop Sci. 2010;15:699–704.

    Article  PubMed  Google Scholar 

  8. Wiemann 4th JM, Tryon C, Szalay EA. Physeal stapling versus 8-plate hemiepiphysiodesis for guided correction of angular deformity about the knee. J Pediatr Orthop. 2009;29:481–5.

    Article  PubMed  Google Scholar 

  9. Gottliebsen M, Rahbek O, Hvid I, Davidsen M, Hellfritzsch MB, Møller-Madsen B. Hemiepiphysiodesis: similar treatment time for tension-band plating and for stapling: a randomized clinical trial on guided growth for idiopathic genu valgum. Acta Orthop. 2013;84:202–6.

    Article  PubMed Central  PubMed  Google Scholar 

  10. Burghardt RD, Herzenberg JE, Standard SC, Paley D. Temporary hemiepiphyseal arrest using a screw and plate device to treat knee and ankle deformities in children: a preliminary report. J Child Orthop. 2008;2:187–97.

    Article  PubMed Central  PubMed  Google Scholar 

  11. Burghardt R, Specht S, Herzenberg J. Mechanical failures of eight plate guided growth system for temporary hemiepiphysiodesis. J Pediatr Orthop. 2010;30:594–7.

    Article  PubMed  Google Scholar 

  12. Oto M, Yılmaz G, Bowen JR, Thacker M, Kruse R. Adolescent Blount disease in obese children treated by eight-plate hemiepiphysiodesis. Eklem Hastalik Cerrahisi. 2012;23:20–4.

    PubMed  Google Scholar 

  13. Schroerlucke S, Bertrand S, Clapp J, Bundy J, Gregg FO. Failure of Orthofix eight-Plate for the treatment of Blount disease. J Pediatr Orthop. 2009;29:57–60.

    Article  PubMed  Google Scholar 

  14. Aykut US, Yazici M, Kandemir U, Gedikoglu G, Aksoy MC, Cil A, et al. The effect of temporary hemiepiphyseal stapling on the growth plate: a radiologic and immunohistochemical study in rabbits. J Pediatr Orthop. 2005;25:336–41.

    Article  PubMed  Google Scholar 

  15. Kanellopoulos AD, Mavrogenis AF, Dovris D, Vlasis K, Burghart R, Soucacos PN, et al. Temporary hemiepiphysiodesis with blount staples and eight-plates in pigs. Orthopedics. 2011;34:263.

    Google Scholar 

  16. Ahmed R, Faisal NH, Paradowska AM, Fitzpatrick ME, Khor KA. Neutron diffraction residual strain measurements in nanostructured hydroxyapatite coatings for orthopaedic implants. J Mech Behav Biomed Mater. 2011;4:2043–54.

    Article  CAS  PubMed  Google Scholar 

  17. Schuh A, Zeller C, Holzwarth U, Kachler W, Wilcke G, Zeiler G, et al. Deep rolling of titanium rods for application in modular total hip arthroplasty. J Biomed Mater Res B Appl Biomater. 2007;81:330–5.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Li Zhao.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

ZW carried out the animal model and histological staining, participated in the design of the study, and drafted the manuscript. DZ helped to make the animal model and performed the statistical analysis. LZ conceived of the study, participated in its design, and helped to draft the manuscript. HL and JL carried out the measurement and analysis of image. DP helped to draft the manuscript and interpreted the data. FM carried out the measured and analysis of residual stress. All authors significantly contributed to the manuscript. All authors have read and approved the final manuscript.

Financial support

This work was financially supported by a grant from the foundation for “985 project” of Shanghai Jiao Tong University School of Medicine (XH0806).

Rights and permissions

Open Access  This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.

The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

To view a copy of this licence, visit https://creativecommons.org/licenses/by/4.0/.

The Creative Commons Public Domain Dedication waiver (https://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wu, Z., Zhao, D., Zhao, L. et al. A comparison between a hinged plate and screw system and a conventional tension-band plate and screw system used for correction of an angular deformity of the lower limb: an animal study. J Orthop Surg Res 10, 57 (2015). https://doi.org/10.1186/s13018-015-0198-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13018-015-0198-4

Keywords